Applications
  • Easydrop/Eurodrop is a single-use disposable flow regulator which allows highly accurate control of flow rate with IV solutions. It is a low-cost alternative to highly expensive IV pump systems (with special IV sets).

  • Easydrop/Eurodrop is easy to use with gravity IV sets and requires "two hands" safety operation for any adjustment.

  • The Easydrop/Eurodrop operating principle is the "triple labyrinth," the GVS patented technology that ensures a very accurate flow rate and high stability in delivery throughout the entire infusion session.

  • No electrical power is required. Easydrop/Eurodrop has a very light weight and modern look.

  • These attributes also make it ideal for home infusion.

  • Easydrop/Eurodrop can be used in a standard IV line or provided in an extension set.

Specifications
Biosafety
  • Materials comply with USP class VI-121°C test and ISO 10993

Dimensions
  • Diameter x Height: 32.00 x 34.20mm
  • Weight: 9.00g

Operating range
  • 20 to 250ml/h*

Inlet/Outlet connectors
  • Standard IV tubing connectors (Tube 2.7 x 3.5 or 3.0 x 4.1mm �?Please ask for drawing)

Materials of construction
  • Housing: White ABS (also available in other colours or materials)
  • Gasket: SEBS (also available in other materials)

Max operating pressure
  • 0.5 bar (7psi) is static condition, gravity set

Max storage and operating temperature
  • 40°C (104°F) indoor operation

Calibrated scale
  • Easydrop/Eurodrop is available in single and double scales.
  • Easydrop/Eurodrop single scale has a general purpose scale that can be used for any kind of solution.
  • Easydrop/Eurodrop double scale has one scale for liquid lower than 10% density (light solutions) and a second scale for liquid with higher density (40%).

Flow rate tolerance
  • range 5-15 ml/h not defined
  • range 20ml/h tolerance on flow rate -10 / + 50%
  • range 21-40ml/h tolerance on flow rate -10 / + 30%
  • range 41-250ml/h tolerance on flow rate -10 / + 20%

Stability of flow rate
  • 10% flow rate fluctuation during 24 hours of infusion (tested with NaCl, 0.9% solution from glass bottles).

Pyrogenicity
  • <0.25 EU/ml using the LAL test method

Sterilization Compatibility
  • Suitable for EtO (44°C / 111 F max.) and gamma (25kGy). Higher temperature can cause a reduction° in performance.